- Flame ionization detector (FID) module designed for highly sensitive detection of organic compounds
- Linear dynamic range greater than 107(±10%) enables accurate quantification of analytes across a broad concentration range
- High sensitivity with minimum detectable level (MDL) less than 1.2 pg C/s and sensitivity greater than 0.03 Coulombs/gC
- Small cell volume and rapid acquisition rate (up to 600 Hz) support both standard and fast GC applications
- Maximum operating temperature up to 450°C
- Integrated electronic gas control (IEC) for precise regulation of air (0 to 500 mL/min), hydrogen (0 to 100 mL/min), and makeup gas (N2 or He, 0 to 50 mL/min)
- Flameout detection with automatic re-ignition improves operational reliability and reduces downtime
- Optimized for capillary columns and compatible with 1/8 in. and 1/16 in. packed columns
- Modular iConnect design allows tool-free installation and replacement in minutes, enabling rapid system reconfiguration
Tekniske data
Tekniske data
| Til brug med (udstyr) | TRACE 1300/1600 Series GC |